Using finite element method comparative analyzes of the stress-strain distribution of the single-rod and double-rod instrumentation for anterior «bone on bone» spinal fusion was done. It was revealed that the most loaded zones of the «thoracic spine-implant system» are rods and anterior parts of the vertebral bodies at the apex of the kyphosis. Close contact of the vertebral bodies in instrumented spine spares physiologic loading of the spine.

Downloads

Download data is not yet available.

References

1. Исследование напряженно-деформированного состояния системы «позвоночник-фиксирующее устройство» при различных вариантах фиксации / А. А. Мезенцев, Д.Е. Петренко, З.М. Мителева [и др.] // Медицина и… – 2006. – №3(14). – С.14-19.

2. Brodner W. Short Segment Bone-on-Bone Instrumentation for Single Curve Idiopathic Scoliosis / W. Brodner, W.M. Yue, H. B. Moller, R.W. Gaines // Spine. – 2007. – Vol. 28. – № 20S. – Р. 224-233

3. Кнетс И.В. Деформирование и разрушение твердых биологических тканей / И.В. Кнетс, Г.О. Пфафрод, Ю.Ж. Саулгозис. – Рига: Зинатне, 1980. – 320 с.

4. Зенкевич О.К. Метод конечных элементов в технике / О.К. Зенкевич. – М: Мир, 1978. – 519 с.

5. Ashman RB Mechanical testing of spinal instrumentation /Ashman RB, Birch JG, Bone LB// CORR. – 1988. – № 227. – P.113-125.